[The main components]: Medical hemostasis is composed of cuffs, one -way valve, inflatable ball, inflatable pipe, and pressure gauge.
[Product Performance]: Stability requirements: It can withstand 80kPa (600mmHg) pressure without falling off or rupturing; Air-tightness requirements:the pressure at 60kPa lasts for 30 minutes and decreases no more than 3kPa. Pressure gauge requirements: Pressure scale range: 0 ~ 100kPa (0 ~ 750mmHg); the minimum score of the pressure gauge is 2KPa and 20mmHg. Pressure reading difference is: ± 10%. [Scope of application]: For clinical venous puncture or bleeding of limb bleeding.
[Instructions for use]:
1. Select the proper cuff size. After the corresponding parts are wrapped,
the inflatable ball is connected to the hemostatic sleeve strap.
2. Screw the deflating knob in the clockwise direction, press the
inflatable ball to the appropriate pressure.
3. Lift the medical hemostatic band : Screw the knob in anticlockwise
direction to decompress, and the cuff is unbuttoned.